A simple and rapid method for simultaneous determination of mycophenolic acid (MPA) and its glucuronide (MPAG) in plasma using high-performance liquid chromatography (HPLC) with UV detection is described. MPA is an immunosuppressant used in kidney, liver and heart transplantation to prevent organ rejection. Owing to MPA’s narrow therapeutic window and considerable variability within and between patients, the routine monitoring of MPA concentrations is suggested.
by C. Misch and Prof. P. Tang PhD
Background
Mycophenolate mofetil (MMF) and enteric-coated mycophenolate sodium (EC-MPS) are widely used to prevent organ rejection after organ transplantation. Following administration, both prodrugs are rapidly hydrolysed to mycophenolic acid (MPA), the active immunosuppressant. MPA is able to suppress the synthesis of guanosine nucleotides in T and B lymphocytes, principally via noncompetitive, selective and reversible inhibition of inosine monophosphate dehydrogenase. MPA is primarily metabolized by the uridine diphosphate glucuronyl transferase to an inactive glucuronide (MPAG), which is transported from liver into bile. Biliary MPAG then enters the gastrointestinal (GI) tract, where it is converted back to MPA, which is then recycled into the bloodstream via the enterohepatic circulation pathway. Several studies have documented that variation in MPA plasma concentrations are unpredictable and variability in plasma concentrations of MPA within and between individuals are high [1–4]. The highly variable set of patient situations on MPA therapies can cause variable risk for adverse effects such as hematologic and GI toxicity. Therapeutic drug monitoring (TDM) of MPA and MPAG can aid clinicians develop personalized therapy strategies to avoid toxicity and maintain efficacy.
For measuring MPA and MPAG concentrations in biological samples, high-performance liquid chromatography (HPLC) has been the primary technique. Scrutinizing all reported technologies, mass spectrometry is generally superior in sensitivity, selectivity and specificity to other detectors. However, the purchase, maintenance and running costs of mass spectrometry are high. From an economic standpoint, HPLC/UV methods [5–10] allow cost-effective assay while provide adequate sensitivity, selectivity and specificity for measuring clinically relevant concentrations of MPA (0.5–5 μg/mL) and MPAG (5–100 µg/mL). The intent of this application was to develop a simple and rapid HPLC/UV method for the determination of MPA and MPAG concentrations in plasma.
Experimental details
Apparatus and materials
The instrument and analytical conditions are listed in Table 1. MPA and internal standard clonazepam were obtained from Sigma (St. Louis, MO). MPAG was from TRC (Toronto Research Chemicals). All other chemicals used were analytical grade or HPLC grade. Separate stock solutions of clonazepam, MPA and MPAG were prepared by accurately weighing and dissolving it in an appropriate amount of methanol.
Calibration/sample preparation
For constructing calibration curves, the concentration ranges of MPA and MPAG were set to 0.1–20 and 1–200 µg/mL, respectively. To 0.1 mL of blank plasma, 0.1 mL each of clonazepam, MPA, MPAG and methanol were added; the mixture was vortex-mixed for 1 min. After centrifugation for 10 min at 10 000 rpm, the supernatant was transferred to an autosampler vial. To 0.1 mL of patient plasma, 0.1 mL clonazepam and 0.3 mL methanol were added and processed as stated above
Results and discussion
Chromatographic separation
A typical chromatogram is presented in Figure 1. These compounds resolved without any overlapping of their peaks or ambiguity in identification. All compounds were eluted within 14 min. No interference was observed in patient samples containing endogenous matrix components, metabolites, xenobiotics and concomitant medication (see Table 2).
Linearity
Good linearities (1/x weighted) were obtained for MPA and MPAG with coefficient of determination (r2) values >0.990 from 0.1 to 20 µg/mL (for MPA) or 1 to 200 µg/mL (for MPAG). The percentage deviation was <15%.
Method validation
Method accuracy and precision data are presented in Table 3. Overall the percentage recovery of MPA and MPAG ranged from 93 to 105%, indicating the consistent, precise, and reproducible extraction efficiency of the method. Both within-run (n=6) and between-run (n=30) precisions were <9%.
Comparison between two HPLC-UV methods
Figure 2a and 2b illustrate comparisons between the current method and reference method. The reference method was also based on a HPLC-UV procedure. The correlation between the two methods was good; the linear regression statistics indicated both r2 values >0.990 (P<0.0001). The linear regression equation for MPA correlation was y = 1.018 x + 0.031 with a standard error value of 0.24; where y, the current method and x, the reference method. The linear regression equation for MPAG correlation was y = 0.984 x − 0.292 with a standard error value of 5.08.
MPA and MPAG concentrations in plasma
Figure 3 illustrates considerable variability of MPA and MPAG concentrations in patient plasma. MPA concentrations ranged from 0.3 to143 µg/mL; MPAG concentrations ranged from 1.2 to 457 µg/mL; MPAG : MPA mole ratio ranged from 0.5 to 186. The mean values for MPA, MPAG and MPAG : MPA were 9.5 µg/mL, 62.3 µg/mL and 13.5, respectively. Clearly, this assay can aid clinicians develop personalized therapy strategies to avoid toxicity and maintain efficacy.
Conclusion
This method includes single dilution step, protein precipitation, ultracentrifugation and gradient chromatography. Sample preparation is rapid and efficient. This method avoids the use of more complex liquid–liquid extraction or solid-phase extraction procedure, which substantially decreases set-up time. This method has been applied to measure MPA and MPAG concentrations in plasma for pharmacokinetic studies and for monitoring clinical use of MPA prodrugs.
References
1. Bullingham RE, Nicholls AJ, Kamm BR. Clinical pharmacokinetics of mycophenolate mofetil. Clin Pharmacokin. 1998; 34: 429–455.
2. Shaw LM, Korecka M, et al. Mycophenolic acid pharmacodynamics and pharmacokinetics provide a basis for rational monitoring strategies. Am J Transplant. 2003; 3: 534–542.
3. Staatz CE, Tett SE. Clinical pharmacokinetics and pharmacodynamics of mycophenolate in solid organ transplant recipients. Clin Pharmacokinet. 2007; 46:13–58.
4. Cattaneo D, Baldelli S, Perico N. Pharmacogenetics of immunosuppressants: progress, pitfalls and promises. Am J Transplant. 2008; 8: 1374–1383.
5. Indjova D, Kassabova L, Svinarov D. Simultaneous determination of mycophenolic acid and its phenolic glucuronide in human plasma using an isocratic high-performance liquid chromatography procedure. J Chromatogr B Analyt Technol Biomed Life Sci. 2005; 817: 327–330.
6. Patel CG, Akhlaghi F. High-performance liquid chromatography method for the determination of mycophenolic acid and its acyl and phenol glucuronide metabolites in human plasma. Ther Drug Monit. 2006; 28: 116–122.
7. Bahrami G, Mohammadi B. An isocratic high performance liquid chromatographic method for quantification of mycophenolic acid and its glucuronide metabolite in human serum using liquid-liquid extraction: application to human pharmacokinetic studies. Clini Chim Acta. 2006; 370: 185–190.
8. Mino Y, Naito T, et al. Simultaneous determination of mycophenolic acid and its glucuronides in human plasma using isocratic ion pair high-performance liquid chromatography. J Pharm Biomed Anal. 2008; 46: 603–608.
9. Watson DG, Araya FG, et al. Development of a high pressure liquid chromatography method for the determination of mycophenolic acid and its glucuronide metabolite in small volumes of plasma from paediatric patients. J Pharm Biomed Anal. 2004; 35: 87–92.
10. Westley IS, Sallustio BC, Morris RG. Validation of a high-performance liquid chromatography method for the measurement of mycophenolic acid and its glucuronide metabolites in plasma. Clin Biochem. 2005; 38: 824–829.
The authors
Catherine Misch MLT and Peter Tang* PhD
Department of Pathology and Laboratory Medicine, Cincinnati Children’s Hospital Medical Center, University of Cincinnati College of Medicine, Cincinnati, Ohio, USA
*Corresponding author
E-mail: peter.tang@cchmc.org
A comparative analysis of serotyping and mass spectrometry (MS) methods for the determination of the flagellar type (H type) of clinically isolated Escherichia coli has been performed. In this analysis, it was shown that determination of the correct H type of a clinical E. coli strain was better achieved by MS than serotyping. Whole genome sequencing was used for the validation of this analysis.
by M. Chan, Dr H. Chui, D. Hernandez, Dr G. Wang and Dr K. Cheng
Background
During outbreaks of disease caused by pathogenic Escherichia coli, it is important to be able to identify the precise E. coli strain involved in order to track and prevent the spread of infection. Typically, the identification of E. coli strains has been based on the serotype of the cell surface antigens such as the lipopolysaccharide O antigen and the flagellar H antigen. Even though serotyping is viewed as the ‘gold standard’ for O- and H-type determination, this technique does have its downfalls. These conventional serotyping methods are based on antisera, which makes procedures costly and laborious to perform because of the variable quality of antibody preparations and the number of antibody agglutination reactions needed to assign a final classification [1, 2, 3]. Also, when bacterial cells do not generate lipopolysaccharide on the surface, the cultured colonies become ‘rough strains’. This makes both O- and H-antigen identification by antibody-based agglutination problematic despite the cellular motility and presence of the flagella H-antigen structure [1, 3, 4]. In addition, flagellar serotyping needs the induction of motility, which can take up to two weeks, and so does not result in the fast identification required in an E. coli outbreak situation [5].
A promising technique for E. coli H-type identification is the use of a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method, termed ‘MS-H’. In our own analytical assays, we harvest, enrich and digest the flagella proteins. LC-MS/MS uses liquid chromatography to separate flagellar fragments after trypsin digestion, and mass spectrometry to analyse and determine the protein sequences of these fragments. The MS results are then analysed against a curated database of H antigen variants to determine the H type. Compared with serotyping, MS-H has a high throughput, requires less labour, needs less time to perform, and provides sequence-level information.
Comparative H typing of E. coli with serotyping and MS-H
Using reference strains of all the different forms of E. coli flagella, it was shown that all 53 different types of H antigen can be determined through MS-H. Table 1 details the comparisons between H serotyping and MS-H [1]. It is important to note that both methods can reach 100% sensitivity and specificity for H-type determination. However, MS-H also provides sequence-level identification of the H type. This is important because this provides more reliable results, while antisera serotyping only provides a visual conformation, making there more of a chance for subjective and inaccurate determination of the H type.
Flagella motility induction is commonly used in serotyping. This increases the time-length of the procedure and requires more hands-on involvement [5]. In comparison, MS-H uses motility induction much less frequently, which makes MS-H identification of E. coli H type quicker and less labour-intensive.
Clinically isolated strains of E. coli that had already had their H types determined by serotyping were supplied from three provinces within Canada. A preliminary H-type analysis of the same E. coli isolates was performed by MS-H [5]. The workflow that was completed to analyse the H type of these strains is detailed in Figure 1.
Things to note within the workflow are that vortexing the suspended culture in Step 2 was to shear the flagella away from the E. coli. Also in Step 2, the flagella were trapped onto a filter membrane. This serves as an isolation step to separate the flagella from the supernatant as well as an enrichment process to concentrate the flagella. We then compared the flagellar sequence obtained by MS to a curated database formed from NCBI flagella protein sequence entries. A curated database was necessary because a public database can be problematic because of size, specificity and inconsistent annotation of the data for the specific flagellar protein entries. The curated database focuses on the flagellar proteins with fit-for-purpose annotations, thus allowing differentiation between the 53 different H types better than the public database [6].
On comparison of the H types identified by serotyping and MS-H, the majority of the results agreed with one another. However, there were some discrepancies between the two methods. In these cases, whole genome sequencing and polymerase chain reaction (PCR) detection was used to identify the correct H type [5] and these results predominantly agreed with those obtained by MS-H.
Even though PCR as a form of detection for the flagella gene is popular, the accuracy for determining all forms of flagella types through PCR is low and the use a many different primers had to be implemented, making it not ideal for the detection of an unknown H type [5]. Whereas whole genome sequencing may be more expensive then PCR detection, it was much more accurate at determining the correct H-type allele than PCR detection. Thus moving into the confirmational assays for the comparative analysis of MS-H determination and H serotyping, whole genome sequencing was only used to resolve disputed results between the two methods as it is not exclusively representative of a host’s flagella phenotype. On the other hand serotyping and MS-H is representative of the bacteria’s H type. Whole genome sequencing isn’t ideal for identifying clinical strains of E. coli for it is laborious and a long process compared to MS-H which is faster and contains less labour-intensive.
MS-H in clinical laboratories
Having shown that MS-H determined E. coli H type with high accuracy and in a short time frame, the application of H typing through MS-H could become very useful for the identification of unknown E. coli strains in a clinical laboratory. This is especially useful in the identification of an E. coli strain during a disease outbreak [7]. Faster identification of the pathogenic strain of E. coli would in turn also help combat the pathogenic E. coli more quickly.
Without the use of antibodies for agglutination and procedures for motility induction that are required in serotyping, MS-H is much less laborious in comparison. This would greatly benefit the professionals in the medical microbiology and public health laboratories who perform H typing.
Advantages and disadvantages of MS-H
It is significant that MS-H better determined the H type compared to conventional serotyping. As MS-H is faster, less laborious, provides sequence-level identification and has a higher throughput than serotyping, MS-H can be very useful in rapid and accurate identification of E. coli flagellar antigens. This may be a little over-idealistic as LC-MS/MS machines are very expensive and serotyping has been around for a very long time. It might be difficult currently for clinical laboratories to afford a LC-MS/MS machine or to change their workflow to incorporate MS-H as their method of H determination. However, mass spectrometer platforms are becoming more common in clinical laboratories. The uses of matrix-assisted laser desorption/ionization (MALDI) mass spectrometers have shown varying successes in microbiological identification. Also the use of MS isn’t just limited to H typing. MS could extend to the determination of the lipopolysaccharides (O antigen) of E. coli, toxins, and other relevant molecules within the spectrum of the mass spectrometer. This would not only determine the surface antigens of pathogenic E. coli, it would also give a broader profile of a pathogenic E. coli.
Conclusion and future directions
MS has potential to determine the H antigen type of E. coli better than serotyping, as shown in a comparative study between MS-H and serotyping through the use of a mass LC-MS/MS platform. MS-H provides sequence level identification of the flagellum type, whereas serotyping only provides visual agglutination assays for positive results and is therefore more prone to errors such as false positives and misidentification. Also, without the need for antisera, MS-H is less laborious, requires less time, and has a high throughput.
With the completion of the preliminary assay results of the comparative study between MS-H and serotyping in the determination of the flagella type, we are currently working on a country-wide thorough validation of the platform.
References
1. Cheng K, Drebot M, McCrea J, Peterson L, Lee D, McCorrister S, Nickel R, Gerbasi A, Sloan A, Janella D, Van Domselaar G, Beniac D, Booth T, Chui L, Tabor H, Westmacott G, Gilmour M, Wang G. MS-H: A novel proteomic approach to isolate and type the E. Coli H antigen using membrane filtration and liquid chromatography-tandem mass spectrometry (LC-MS/MS). PLoS One 2013; 8(2): 1–12.
2. Tenover FC, Arbeit RD, Goering RV. How to select and interpret molecular strain typing methods for epidemiological studies of bacterial infections: A review for healthcare epidemiologists. Molecular Typing Working Group of the Society for Healthcare Epidemiology of America. Infect Control Hosp Epidemiol. 1997; 18(6): 426–439.
3. Machado J, Grimont F, Grimont PA. Identification of Escherichia coli flagellar types by restriction of the amplified fliC gene. Res Microbiol 2000; 151(7): 535–546.
4. Edwards PR, Ewing WH. Edwards and Ewing’s Identification of Enterobacteriaceae, p536. Elsevier 1986. ISBN 978-0444009817.
5. Cheng K, Sloan A, Peterson L, McCorrister S, Robinson A, Walker M, Drew T, McCrea J, Chui L, Wylie J, Bekal S, Reimer A, Westmacott G, Drebot M, Nadon C. Knox D, Wang G. Comparative study of traditional flagellum serotyping and liquid chromatography-tandem mass spectrometry-based flagellum typing with clinical Escherichia coli isolates. J Clin Microbiol. 2014; 52(6): 2275-2278.
6. Cheng K, Sloan A, McCorrister S, Babiuk S, Bowden TR, Wang G, Knox D. Fit-for-purpose curated database application in mass spectrometry-based targeted protein identification and validation. BMC Res Notes 2014; 7: 444.
7. Cheng K, Sloan A, McCorrister S, Peterson L, Chui H, Drebot M, Nadon Celine, Knox D, Wang G. Quality evaluation of LC-MS/MS-based E. coli H antigen typing (MS-H) through label-free quantitative data analysis in a clinical sample setup. Proteomics Clin Appl. 2014; 8: 963–970.
The authors
Michael Chan*1,2, Huixia Chui1,3 MD, Drexler Hernandez1,2, Gehua Wang*1 MD and Keding Cheng*1,4 MD
1National Microbiology Laboratory, Public Health Agency of Canada, Winnipeg, MB, Canada
2University of Manitoba, Winipeg, MB, Canada
3Centre of Disease Control and Prevention, Henan Province, PR China
4Department of Human Anatomy and Cell Sciences, Faculty of Medicine, University of Manitoba, Winnipeg, MB, Canada
*Corresponding authors
E-mail: M. Chan, umchanm@myumanitoba.ca; G. Wang, gehua.wang@phac-aspc.gc.ca; K. Cheng, keding.cheng@phac-aspc.gc.ca.
February | March 2025
The leading international magazine for Clinical laboratory Equipment for everyone in the Vitro diagnostics
Beukenlaan 137
5616 VD Eindhoven
The Netherlands
+31 85064 55 82
info@clinlabint.com
PanGlobal Media is not responsible for any error or omission that might occur in the electronic display of product or company data.
This site uses cookies. By continuing to browse the site, you are agreeing to our use of cookies.
Accept settingsHide notification onlyCookie settingsWe may ask you to place cookies on your device. We use cookies to let us know when you visit our websites, how you interact with us, to enrich your user experience and to customise your relationship with our website.
Click on the different sections for more information. You can also change some of your preferences. Please note that blocking some types of cookies may affect your experience on our websites and the services we can provide.
These cookies are strictly necessary to provide you with services available through our website and to use some of its features.
Because these cookies are strictly necessary to provide the website, refusing them will affect the functioning of our site. You can always block or delete cookies by changing your browser settings and block all cookies on this website forcibly. But this will always ask you to accept/refuse cookies when you visit our site again.
We fully respect if you want to refuse cookies, but to avoid asking you each time again to kindly allow us to store a cookie for that purpose. You are always free to unsubscribe or other cookies to get a better experience. If you refuse cookies, we will delete all cookies set in our domain.
We provide you with a list of cookies stored on your computer in our domain, so that you can check what we have stored. For security reasons, we cannot display or modify cookies from other domains. You can check these in your browser's security settings.
.These cookies collect information that is used in aggregate form to help us understand how our website is used or how effective our marketing campaigns are, or to help us customise our website and application for you to improve your experience.
If you do not want us to track your visit to our site, you can disable this in your browser here:
.
We also use various external services such as Google Webfonts, Google Maps and external video providers. Since these providers may collect personal data such as your IP address, you can block them here. Please note that this may significantly reduce the functionality and appearance of our site. Changes will only be effective once you reload the page
Google Webfont Settings:
Google Maps Settings:
Google reCaptcha settings:
Vimeo and Youtube videos embedding:
.U kunt meer lezen over onze cookies en privacy-instellingen op onze Privacybeleid-pagina.
Privacy policy